Channel Calibration Procedure

Instructions for Calibrating the Datatron Channels

 

Note: Alarm Functionality is Temporarily Suspended for the Channel that is currently displayed on the Calibration Page.

Procedure

KEY : RO = Remote operator @ sensors, CO = Computer operator

 

I.  RTD Sensor and Temperature Transmitter Calibration

 

II.  Humidity Transmitter Calibration

 

III.  Pressure Transmitter Calibration

    1. RO : Identify the channel to be calibrated and tell CO channel number

    2. CO : Bring channel into calibration screen and inform RO.

    3. RO : Disconnect wires from "+" and "-" terminals. Observe which wire color goes to "+" and "-". Inform CO that power is removed.

    4. CO : Observe that the transmitter output value has changed, verifying the power has been disconnected.

    5. RO : Attach 4-20mA calibrator to wires removed from transmitter circuit board. BE SURE TO OBSERVE POLARITY. Set unit output to 4 mA position.

    6. RO : Inform CO that unit is in "Zero" position.

    7. CO : Click mouse on "Calibrate" button on left side of window.

    8. CO : Enter the minimum value of the transmitter range.

    9. CO : Click on "Ok" button to start. Computer takes readings and displays new window requesting "Span" setting. Inform RO to go to 20mA.

    10. RO : Set box to output 20mA and inform CO.

    11. CO : Enter the maximum value of the transmitter range and click on "Ok" button.

    12. CO : Computer takes readings and displays a new window requesting that sensor to be re-attached. Inform RO to attach wires back onto transmitter.

    13. RO : Connect wires to "+" & "-" terminals and connect the pressure tubes on the transmitter to each other. Inform the CO.

    14. CO : Enter 0.0 for the reading and click on "Ok". The system takes the reading and displays the new calibration factors to the right of the Initial factors. Verify that the new scale factors are within the acceptable limits. Final calibration results must be within these ranges. If not, calibration must be repeated. If the repeated calibrations do not fall within the acceptable limits for the sensor range, replace the transmitter and sensor and re-calibrate.

    15. RO : Reconnect the pressure tubing, close transmitter box and wait for CO instructions.

    16. CO : Click on "Save" to save new data, click on "Yes" for making changes permanent.

    17. CO : Once complete, click on "Print" to print out calibration data. Sign and date the report.

    18. RO : Close transmitter box, apply a calibration sticker that has been initialed and dated to the cover of the sensor.  Proceed to next unit.

 

IV. CO2 Transmitter Calibration

    1. RO : Identify the channel to be calibrated and tell CO channel number.

    2. CO : Bring channel into calibration screen and inform RO.

    3. RO : Disconnect the transmitter and inform CO.

    4. CO : Observe that the transmitter output changes, verifying the disconnection of the power.

    5. RO : Attach the 4-20mA Calibrator to the wires removed from the transmitter.  BE SURE TO OBSERVE POLARITY.  Set the box output to 4mA position.

    6. RO : Inform CO that unit is in "Zero" position.

    7. CO : Click mouse on "Calibrate" button on left side of window.

    8. CO : Enter 0.0 for the low value.

    9. CO : Click on "Ok" button to start. Computer takes readings and displays new window requesting "Span" setting. Inform RO to go to 20mA.

    10. RO : Set the output to the 20mA position and inform CO.

    11. CO : Enter 20.0 for span value and click on "Ok" button.

    12. CO : Computer takes readings and displays a new window requesting that sensor to be re-attached.

    13. RO : Connect the 5% CO2 gas bottle to the sensor and set the flow rate to 100 cc/min.

    14. CO : Enter 5.0 for the reading and click on "Ok". The system takes the reading and displays the new calibration factors to the right of the Initial factors. Verify that the new scale factors are within the acceptable limits. Final calibration results must be within these ranges. If not, calibration must be repeated. If the repeated calibrations do not fall within the acceptable limits for the sensor range, replace the transmitter and sensor and re-calibrate.

    15. CO : Click on "Save" to save new data, click on "Yes" for making changes permanent if Ok.

    16. CO : Once complete, click on "Print" to print out calibration data. Sign and date the report.

    17. RO : Stop gas flow rate and remove cylinder and sensor gas tube.

    18. RO : Close transmitter box, apply a calibration sticker that has been initialed and dated to the cover of the sensor.  Proceed to next unit.

 

V. Calibration Results

    1. Final calibration results must be within these ranges. If not, calibration must be repeated. If the repeated calibrations do not fall within the acceptable limits for the sensor range, replace the transmitter and sensor and re-calibrate.  All calibration reports must be signed and dated.  The calibration reports must be given to the Director of Engineering for final review.  Once the reports have been reviewed, a Certificate of Calibration will be issued to the user by the Engineering Department.

 

VI. Calibration Acceptance Limits

    1. Please Contact R&D Scientific for a current list of Acceptance Limits

 

VII. EQUIPMENT REQUIRED

    1. Please Contact R&D Scientific for a current list of Required Calibration Equipment for your Datatron System